Edit points, Edit points -11 – RIGOL DG5000 Series User Manual
Page 71
Chapter 3 Arbitrary Waveform Output
RIGOL
User’s Guide for DG5000
3-11
Edit Points
The following example introduces how to create an arbitrary waveform which meets
with the following conditions by using “Edit Points”.
Operating Procedures:
1. Press Arb, use
to open the 2/2 menu page, and then press Create New
Cycle Period, use the numeric keyboard to input “12” and select the unit
“μsec” in the pop-up menu.
2. Press High VLimit, use the numeric keyboard to input “4” and select the unit “V”
in the pop-up menu.
3. Press Low VLimit, use the numeric keyboard to input “-2” and select the unit “V”
in the pop-up menu.
4. Press Init #Points, use the numeric keyboard to input “4” and then press OK.
Note that a level line of -2 V appears.
5. Press Interp to select “Linear”.
6. Press Edit Points to enter the “Edit Point” interface.
1) Press PointID to define the first point (default time is 0). Press Voltage and
Parameter
Value
Cycle Period
12 μs
High VLimit
4 V
Low VLimit
-2 V
Interpolation
Linear
Point 1
0 s, 0 V
Point 2
4 μs, 4 V
Point 3
8 μs, 0 V
Point 4
10 μs, -2 V