6 - 90
Chapter 6 System Setup
Running the DATRACK part program
Run the DATRACK part program to establish and save the coordinate system
of the rack. There are eight DATRACK programs stored in the c:\qc5000\Probe
Rack folder. Select the appropriate program for the probe rack installed.
4 port rack, X Positive
DATRACK4XP.5pa
4 port rack, Y Positive
DATRACK4YP.5pa
4 port rack, X Negative
DATRACK4XN.5pa
4 port rack, Y Negative
DATRACK4YN.5pa
6 port rack, X Positive
DATRACK6XP.5pa
6 port rack, Y Positive
DATRACK6YP.5pa
6 port rack, X Negative
DATRACK6XN.5pa
6 port rack, Y Negative
DATRACK6YN.5pa
NOTE
Use a 2mm probe tip with a 20 mm stylus when running the
DATRACK program.
To run the DATRACK program:
1
Click File/Open to display the Part Open file dialog box, then navigate to
C:\QC5000\ProbeRack, select the desired DATRACK program and click Open.
2
Click the New Run button on the program toolbar to execute the selected
DATRACK program.
3
Follow the on-screen prompts to complete the
DATRACK program.