6 - 29

Quadra-Chek

®

5000

Step 3
True magnification must be enabled. If the True magnification box is not checked,
click on True magnification and then click Show.

Step 4
Click OK to close the Customize screen, and then return to the Tools/Options/
VED tab
.

Step 5
Bring a standard artifact into sharp focus in the video Window’s field of view.
The Metronics calibration slide is an excellent standard to use for this proce-
dure.

Step 6
Measure the distance between 2 of the
standard’s marks on the computer
screen using a scale or ruler. Measure
as great a distance as possible.

Step 7
Calculate the measured magnification
by dividing the measured distance
(Step 6) by the actual distance shown
on the standard.

Measured magnification = Measured / Actual

Step 8
Calculate the pixel size factor by dividing the measured magnification by the
magnification value currently displayed in the right corner of the status bar.

Pixel size factor = Measured Magnification / Status bar value

Step 9
Enter the pixel size factor into the X pixel
size... field
.

NOTE

The value shown here is only an example.