HEIDENHAIN IK 5394-3D User Manual
Page 514

6 - 64
Chapter 6 System Setup
Level-based button functions
Many button functions are invoked only once when the button is pressed. Others
can be configured to be momentary (level-based) or to toggle functions on or
off. Momentary functions are active only while the button is pressed and held.
Toggle functions are activated when the button is pressed and deactivated when
the button is pressed again.
Button functions that are invoked only once
Button functions that are invoked only once when the button is pressed in-
clude:
Measure:
OK
Measure:
Enter pt
Measure:
Remove last
Measure:
Cancel
View from probe
Go to here
Pause program
Run program
NOTE
None of the eight button functions listed above can be
configured as momentary (level-based).
Button functions that can be momentary or toggled
Buttons that are active while pressed and held, or toggled on and off include:
Axis lock
Speed toggle
Motors off
Swap Z joystick
Swap Z digital positioner
Swap Q joystick
Configure any of the six button functions listed above as level-based by check-
ing the button’s Level-based box.
Button functions
Descriptions of the functions are provided in the following paragraphs:
No assignment
Pressing the button causes no action