5000
Pixel calibration
Use the pixel calibration to calibrate camera pixels to the glass master standard.
Perform a pixel calibration for each magnification.
To perform a pixel calibration
Step 1
Position the largest circle on the standard in the center of the field of view as
shown.
Step 2
Click on the new probe button on the VED toolbar.