HEIDENHAIN IK 5394-3D User Manual
Page 478

6 - 28
Chapter 6 System Setup
Calibrating the display of magnification value
The video window image magnification is displayed in the bottom-right corner
of the QC5000 main screen.
The image magnification depends on the optical characteristics of the system
and must be found by performing the measurement procedure provided below.
Calibration procedure
Step 1
Verify that the X pixel size... field is set to
1.000
Step 2
Click Tools, Customize and then click the Status bar tab.
This manual is related to the following products: