6 - 87

Quadra-Chek

®

5000

Probe rack installation

Set machine zero each time you launch the QC-5000. A repeatable machine
zero position is required to find the probe rack.

Follow the instructions of the probe rack manufacturer to install the probe rack
on your system. Once the probe rack is mounted and it’s position entered in the
QC-5000, it cannot be moved.

NOTE

Mount the probe rack so the current probe can reach each of

the ports, and so there is no probe rack movement in any

direction.

Probe rack orientation

The four orientations accepted by the QC-5000 are shown below.

+X Orientation

Along the back edge of the table with the docking side of the rack facing in the
+X direction.

+Y Orientation

Along the left edge of the table with the docking side of the rack facing in the +Y
direction.

-X Orientation

Along the front edge of the table with the docking side of the rack facing in the
-X direction.

-Y Orientation

Along the right edge of the table with the docking side of the rack facing in the
-Y direction.