HEIDENHAIN IK 5394-3D User Manual
Page 179

3-7
Quadra-Chek
®
5000
VED Calibrations
The following calibrations are required before using VED with the QC-5000:
•
Auto focus teach (each magnification)
•
Edge calibration (each VED probe tool)
•
Pixel calibration (each magnification)
•
Camera calibration (each camera)
•
Parcentricity calibration (each magnification)
Use the glass master standard shipped with the QC-5000 to perform these cali-
brations.
NOTE
Perform all VED calibrations from the highest magnification to
the lowest magnification. For example, if there are three
magnifications (low, medium, and high) perform calibrations
on high first.
NOTE
Ignore calibration warnings as shown during the calibration
process. Click the hide warning button and continue the
calibration. The QC-5000 displays this message to warn users to
complete calibrations before performing inspections.
To perform an auto focus teach
Step 1
Highlight the desired magnification as shown.