3-18

Chapter 3 Probes

Step 8
Click on the resolution tab and then click on teach.

Step 9
Enter the measured diameter of the standard circle and click continue as shown.

NOTE

Do not enter the diameter as measured with the QC-5000. Use

the diameter found on the standard’s case.

Step 10
Click enter pt in the dialog box as shown.

NOTE

At least 90 points are required to perform a pixel calibration.