HEIDENHAIN IK 5394-3D User Manual
Page 190

3-18
Chapter 3 Probes
Step 8
Click on the resolution tab and then click on teach.
Step 9
Enter the measured diameter of the standard circle and click continue as shown.
NOTE
Do not enter the diameter as measured with the QC-5000. Use
the diameter found on the standard’s case.
Step 10
Click enter pt in the dialog box as shown.
NOTE
At least 90 points are required to perform a pixel calibration.