6 - 89

Quadra-Chek

®

5000

Dropping in Z

The distance the Z axis moves into the rack during a probe drop.

Pickup in Z

The distance the Z axis moves from the docking port after probe pickup.

Release in Z

The distance the Z axis moves from the docking port during a probe drop.

Approach in Z

The distance the Z axis moves into the rack during a probe pickup.

Entering port position offset

The offset distance along the axis of the rack between the center of the docking
module and the center of the probe tip used for calibration. Enter port position
offsets in the QC-5000 as follows:

Rack orientation is X Positive

Enter the X value

Rack orientation is X Negative

Enter the negative of the X value

Rack orientation is Y Positive

Enter the Y value

Rack orientation is Y Negative

Enter the negative of the Y value

Checking the probe rack orientation

The probe rack must be aligned within 1 degree of the nominal X or Y axis
orientation to ensure reliable probe rack operation.

NOTE

When error correction is enabled, the error correction (NLEC or

SLEC) data entry must be performed before calibrating the probe

rack.

1

Lock Z-axis motion and measure a line along the front of the probe rack.

2

Re-adjust the orientation of the probe rack if the angle of the line is greater

than 1.0 degree from the nominal orientation.