The general tab contains the following checkboxes:

Use machine reference frame

Empty feature list before running

Use probe teach values as recorded

Do not report any errors during program running

Prevent program from being modified by anyone but the supervisor

On a measurement failure goto label

Randomly measure this percent of guide features

Use machine reference frame
Check to use machine zero at the start of a program. Uncheck to use part zero.

Empty feature list before running
Check to delete all features from the features list before running.

Use probe teach values as recorded
Check to use previously recorded probe teaches.

Do not report any errors during program running
Check to ignore all errors while running the program

Prevent program from being modified by anyone but the supervisor
Check to require the supervisor password from users before allowing changes.

On a measurement failure goto label
Check to create a goto step if a measurement fails.

Randomly measure this percent of guide features
Used to measure a random number of features on a part, normally guide fea-
tures are features that are never measured except in this special case where a
percentage of them will be.