Parcentricity calibration

Perform a parcentricity calibration to ensure the optics center of each lens is
properly aligned. All systems with more than one magnification require
parcentricity calibration. Parcentricity calibrations are performed only after all
other calibrations are complete. Use a circle on the standard that will fit the
field of view in all magnifications.

To perform a parcentricity calibration

NOTE

In this demonstration the following magnifications appear in

probe library: High, MedHigh, Medium, MedLow, and Low.

Individual users may have selected different names for the

magnifications stored in probe library. To duplicate this process

use the highest magnification for step 1.

NOTE

Do not move the stage between magnifications while

performing the parcentricity calibration.

Step 1
Highlight the highest magnification as shown and click set current.