Running the Flicker Test

Before starting a test, make sure all user fields are set correctly. Changes to these fields are no
longer possible once a test is in progress. Also make sure the EUT is connected to the AC output
of the PACS unit and the correct AC line voltage is programmed on the AC source.

A test run is started by clicking on the Start button.

During the test execution, the CTS 3.0 Software will continuously evaluate EUT Flicker against the
appropriate IEC limits. The Pass or Fail indication is updated in real-time. If the test fails early in a
long test period, you may elect to abort the test prematurely rather than wait till the end.

6.5

Printing Results

The CTS 3.0 Software can produce a WORD based report. At the end of test, the report button
will be enabled. Clicking on the Report button will generate information needed for a report and
present an option to generate a report in WORD.

6.6

Flicker Replay Mode

The CTS 3.0 Software can be used to replay previously recorded test data. This replay can be
done on the same test system that was used to record the original data or on an altogether
different PC.

In order to replay a test, the data acquired during the test needs to be saved to a CTS native data
file. If the save data option is not enabled in the additional setup window for Flicker, there is no
way to replay a test later. The Pass or Fail result of any run on the actual test system PC is
always logged however. If a different PC is used to play back data, the data files have to be
transferred to this PC first. The test data files are stored in the following directory:

C:\Program Files\California Instruments\CTS\Data_files

6.6.1

To play back test on the actual test system PC, proceed as follows:

1. From the Test Menu, select 'Replay an existing test'. This will bring up a table showing all

tests ever run on this system.

2. Select the test you want to replay by clicking on its row. If data for this test was stored and is

available, the Replay Test button at the bottom will be enabled.

3. Click on the Replay Test button to start the replay process. This will bring up a standard

Harmonics or Flicker test screen.

4. To start the replay process, click on the Start button. The data displayed will be taken from

the stored data files. The replay mode is indicated in the lower right corner on the status bar.

5. The buffer rate and speed at which data is replayed can be controlled using the various

settings in the Options menu. This may be required to adjust to the speed of the PC used.

6. The replay process can be halted at any time by clicking on the Stop button. This will stop the

data at the buffer number shown in the lower left corner.

7. Once stopped, the data can be stepped forward or backward at various rates using the VCR

style buttons in the lower left corner.

8. A test report can be printed from stored data in this mode. The actual date and time the data

was recorded will be shown on the report in addition the report print date and time.

6.6.2

To play back test on a different PC from the test system PC, proceed as follows:

1. If not already done, install a copy of the CTS 3.0 software on the other PC. When launched,

the absence of a data acquisition system will be detected and flagged. This is normal.

2. Transfer the data files to the PC used for replay, either using a network connection or disk.