User Manual

Compliance Test System 3.0

85

Field /
Control

Location

Description

Power
button

Upper left
corner

Click on Power to toggle the power source output voltage, unless
power source is already programmed to 230V. The Flicker module
can be configured to turn off the power after every test, in that case
therefore, the user will have to click Power On before starting the
test.

Start button

Upper left
corner

Starts the flicker test using presently selected parameters. Once a
test is running, parameter cannot be changed. Starting a test also
causes data to be written to the test data file.

Stop button

Upper left
corner

Stop the acquisition process. The Stop button can be used to abort
a test prematurely. The data files will be closed and will only contain
data up to the point at which the test was aborted.

Report Button

Upper left
corner

Used to produce a test report in WORD. A WORD must be
installed to generate the report.

Power Button

Upper left
corner

Can manually toggle the power on/off. A PC to power source
communication is needed before it is enabled.

Frequency

Left Panel

Displays signal frequency

V-rms (volt)

Left Panel

Displays Voltage rms value

Test Start
Time

Left panel

This field always shows the start time of the test in progress. The
test duration is shown in the bottom panel.

Highest d

c

%

Left panel

Displays the highest Relative Steady State voltage change (d

c

) in %

found so far during the test. This represents the difference between
two adjacent steady-state voltages relative to the nominal voltage.
The standard requires that d

c

must be less than or equal to 3 % for

the EUT to PASS.

Highest d

max

% Left panel

Displays the highest Maximum relative voltage change (d

max

) in %.

found so far during the test. This represents the difference between
the maximum and minimum rms values of the voltage change
characteristic relative to the nominal voltage. The standard requires
that d

max

must be less than or equal to 4 % for the EUT to PASS.

Highest d

t

%

Left panel

Displays the highest Relative voltage change characteristic (d

t

).

This value represents the change in rms voltage, relative to the
nominal voltage, as a function of time and between periods when
the voltage is a steady state condition for at least 1 second. The
standard requires that d

t

must be less than 3 % for the EUT to

PASS, although it is permitted to be between 3 and 4 % for less
than 200 msec.

Highest Psti

Left panel

Displays the highest Short Term Flicker value found so far during
the test. Each Short Term Flicker severity is evaluated over period
of 10 minutes. The threshold of irritability is Pst = 1 and this value is
used as the PASS/FAIL limit.

Highest Plt

Left panel

Displays the highest Long Term Flicker value (Plt) found so far
during the test. The Plt period is 120 minutes and is calculated
using successive Psti values. The threshold of irritability for long
term flicker is 0.65 and this value is used as the PASS/FAIL limit.

Test Status

Center
panel

This field provides a visual indication of the test result. A Green field
with Pass indicates the EUT passes the Flicker test, a Red field
with Fail indicates the EUT causes too much flicker. Note that the
condition of this field is affected by the user selected test margin.
To use the actual IEC test limits, the test margin should be set to