Compliance Test System 3.0

135

User Prompts

Some EUT's will require resetting and readjustment when the power dips or is interrupted for a
long period. If this is the case, then set GapDelay0 = -1. This will cause the test sequence to halt
and wait for user intervention. When ready to resume test press “OK”.

14.2.2 IEC413.413 File

[IEC-1000-4-13]

IEC 61000-4-13 section header

Parameters that apply only to Flat Top Curve

Fcurv_lev = 90

Flat Curve Level at 90%

Fcurv_dwe = 120

Flat Curve Dwell time in seconds

Fcurv_pau = 5

Flat Curve Pause in seconds

Parameters that apply only to Over Swing Curve

Oswcurv_dwe = 120

Over Swing Curve Dwell time in seconds

Oswcurv_pau = 5

Over Swing Curve Pause in seconds

Oswcurv3_lev = 6

Over Swing Curve of third harmonic level at 6%

Oswcurv3_phs = 180

Over Swing Curve of third harmonic phase test of 180 degrees

Oswcurv5_lev = 4

Over Swing Curve of fifth harmonic level at 4%

Oswcurv5_phs = 0

Over Swing Curve of fifth harmonic phase test of 0 degree

Parameters that apply to Frequency Sweep

Swefreq1_lev = 3

Sweep Frequency Level of first range at 3%

Swefreq2_lev = 10

Sweep Frequency Level of second range at 10%

Swefreq3_lev = 5

Sweep Frequency Level of third range at 5%

Swefreq4_lev = 5

Sweep Frequency Level of fourth range at 5%

Swefreq5_lev = 2

Sweep Frequency Level of fifth range at 2%

Swefreq_dwe = 300

Sweep Frequency Dwell time in seconds

Swefreq1_stp = 5

Sweep Frequency Step size of first range in Hz

Swefreq2_stp = 10

Sweep Frequency Step size of second range in Hz

Swefreq3_stp = 10

Sweep Frequency Step size of third range in Hz

Swefreq4_stp = 10

Sweep Frequency Step size of fourth range in Hz

Swefreq5_stp = 25

Sweep Frequency Step size of fifth range in Hz

Parameters that apply only to Individual Harmonics

Indharm2_lev = 3

Individual harmonics Level for nth (second) harmonic in %

Indharm2_phs = 0

Individual harmonics Phase for nth (second) harmonic in degree

Indharm_dwe = 5

Individual harmonics Dwell time in second

Indharm_pau = 5

Individual harmonics Pause in second

Parameters that apply only to Interharmonics

Interharm1_lev = 2.5

Interharmonics Level for first range at 2.5%

Interharm2_lev = 5

Interharmonics Level for second range at 5%

Interharm3_lev = 3.5

Interharmonics Level for third range at 3.5%

Interharm4_lev = 2

Interharmonics Level for fourth range at 2%

Interharm5_lev = 1.5

Interharmonics Level for fifth range at 1.5%

Interharm_dwe = 5

Interharmonics Dwell time in second

Interharm_pau = 1

Interharmonics Pause in second