8.4

Test Sequence

The IEC 61000-4-13 standard specifies the following test flow diagram.

Start - Class 1/2

Determine appropriate test

configuration

Perform 8.2.1 Test

"Harmonic combination"

Flatcurve and Overswing

Any functional

anomalies?

No

No

Class 2

required?

'Meister

curve' required?

Any functional

anomalies?

Any functional

anomalies?

Any functional

anomalies?

Test Completed

FAIL

Record Results

Test Completed

PASS

Record Results

Yes

Yes

No

Perform 8.2.2 Test

"Frequency Sweep"

Perform 8.2.4 Test

"Meister Curve"

Perform 8.2.3 Test

"Individual Harmonics/

Interharmonics"

Yes

Yes

No

Yes

Yes

No

No