beautypg.com

Configuration procedure – AMETEK CTS 3.0 User Manual

Page 153

background image

User Manual

Compliance Test System 3.0

149

17.5

Configuration Procedure

1. Start the Calibration module located in the Compliance Test System Program group called

Calibration (See Figure 17-1).

2. Fill in the information for PACS serial #, Power source to be used for the calibration, and your

name or ID. Make all other appropriate selections before clicking on the Calibrate button.

3. Once the Calibrate button is clicked, the software will prompt you to fill in a table recording the

equipment and measurement instruments to be used for this calibration. When this is
completed, the program will provide you with the step-by-step instructions. Follow these
instructions until the Accept button is enabled. If you did not follow the instructions during the
calibration or the software is aborted for some other reason, you may have to redo the entire
calibration. Please note that the calibration data is not saved until the Accept button is pressed.

4. Please note that when calibrating a three-phase PACS, all three channels will be calibrated

together. The software will do all three-voltage channels, all high current range channels, all
mid-range channels, and then all low range channels.

5. When calibrating the voltage channels, you will be asked to provide a 230V/50Hz input.

However, when calibrating the current channels, it is not necessary to keep the voltage input at
the same level. In addition, the calibration on current channels is in the sequence of
40Arms/200App, 8Arms/40App, and 2Arms/10App. The software does not do auto ranging in
the process. This means that even if you apply 2 Arms when you are doing 40Arms/200App
range channel(s), the software will still perform the corresponding calibration. The drawback of
doing so is that the calibration accuracy may be compromised.

6. When the Accept button becomes available and is clicked at the end of calibration, the software

will save the data in text file with the name of “C:\CTS calibration.dat”. In addition, it will
generate “calibrate.log” in the CTS directory and print out a hardcopy of this file automatically.
It will also save the calibration information in a database. Further, it will ask you if you want to
copy the calibration data to floppy. If you answer yes, it will perform all the copying for you
automatically.

7. If you copied the calibration data to the floppy, remove disk from drive and write the PACS serial

number on the floppy disk and keep for backup purposes.