beautypg.com

Calibration, About this chapter – AMETEK CTS 3.0 User Manual

Page 151

background image

User Manual

Compliance Test System 3.0

147

17. Calibration

17.1

About This Chapter

This chapter covers two basic functions: System calibration and system configuration. Single phase
calibration is covered first and additional steps for three phase calibration are also covered. System
configuration is then covered detailing the steps to properly set up the AC power source.

If the CTS system you have includes an AC power source, refer to the separate user manual that
shipped with the AC source for service and calibration information.

17.2

Calibration

The CTS system uses a precision measurement system that requires periodic calibration. The
recommended calibration interval is one year. All CTS systems are factory calibrated before
shipment, so adjustment should not be needed until after the first year of use.

Note: For calibration at the factory, the A/D card must be returned with the PACS.

A special calibration program is included in the CTS 3.0 Software suite to facilitate calibration and
store the calibration constants. This module is called “Calibration” and is available by clicking on
the Compliance Test System menu selection. The program screen is shown on the next page.
Make appropriate selection before clicking on the Calibrate button. Upon Clicking the Calibration
button, you will be given step-by-step instructions to complete the calibration. At the end of a
calibration, the Accept button will be available. Clicking on it will cause the software to save the
calibration data and printout a calibration log file.

17.2.1 Required Calibration Equipment

To perform the CTS measurement system calibration, the following equipment is required:

Equipment

Description

Digital AC Multimeter

Agilent 34401A DMM or equivalent with traceable calibration.

Resistive Load Bank

Avtron K565 or equivalent. The load resistor range must be able to
handle 2 - 30A rms (or higher for PACS-3-75) and be rated for the
power levels used.

Current Shunt

50A Current shunt 0.05 % accuracy or better, e.g. Isotek.

Table 17-1: Required Calibration Equipment