10.4

Waveform Display

The DC output waveform is acquired at regular intervals during the test and can be displayed by
selecting the Waveform Display tab.

10.5

Test Results

There is no way for the CIGUI to determine the outcome of this test as a clear definition of a
failure can not be given. The draft standard called out the following possible test results:

“The�test�results�shall�be�classified�on�the�basis�of�the�operating conditions and functional
specifications of the equipment under test, as in the following, unless different specifications are given
by product committees or product specifications.

a) Normal performance within the specification limits

b) Temporary degradation or loss of function or performance which is self-recoverable.

c) Temporary degradation or loss of function or performance which requires operator intervention or

system reset.

d) Degradation or loss of function which is not recoverable due to damage of equipment

(components) or software, or loss of data.

As a general rule, the test result is positive if the equipment shows its immunity, for the duration of the
application of the test, and at the end of the tests the EUT fulfills the functional requirements
established�in�the�technical�specification.”

The user has to make this determination by observing the unit under test. At the end of a test run,
the following dialog box is displayed. Click on the Help button to display the pass/fail criteria
shown above.

Figure 10-2: IEC 61000-4-17 Waveform Acquisition Window