4-2 Basic Settings and Advanced Settings
4-29
IL-E
4
Setting V
ar
iou
s F
unct
ions
Setting range for analog output scaling
If free range or bank is selected for the analog output scaling state, the upper limit and
lower limit can be set within the range shown below.
IL-S025/IL-030/IL-S065/IL-065/IL-100
IL-300/IL-600
IL-2000
The analog output can be selected from ± 5 V, 1-5 V, 0-5 V or 4-20 mA.
*1 The averaging is as follows after 512 times.
512 times: Sampling rate x 2
1024 times: Sampling rate x 4
2048 times: Sampling rate x 8
4096 times: Sampling rate x 16
Item
Setting range
Default value
Upper limit
-99.999 to 99.999
10.000
Lower limit
-99.999 to 99.999
-10.000
Item
Setting range
Default value
Upper limit
-999.99 to 999.99
100.00
Lower limit
-999.99 to 999.99
-100.00
Item
Setting range
Default value
Upper limit
-9999.9 to 9999.9
1000.0
Lower limit
-9999.9 to 9999.9
-1000.0
Analog Output Accuracy
Voltage output
Current output
Output range
± 5 V
(Full scale 10 V)
4 - 20 mA
(Full scale 16 mA)
Output resistance
100
Ω
-
Maximum load resistance
-
350
Ω
Repeatability
± 1 mV
± 1.5 µA
Accuracy in respect to
display
± 0.05% of F.S.
± 0.25% of F.S.
Temperature
characteristics
± 0.005% of F.S./°C
± 0.01% of F.S./°C
Refresh rate
*1
Same as sensor head sampling rate