4 setting the tolerance setting value, 4 setting the tolerance setting value -10, Setting the tolerance setting value -10 – KEYENCE IL Series User Manual
Page 42: Setting the tolerance setting value
3-10
IL-E
3
Basic Oper
ations
3-4
Setting the Tolerance Setting Value
The tolerance setting value consists of the upper limit value (HIGH side setting value) and
the lower limit value (LOW side setting value). By setting these values, judgments are made
in three levels: when the judgment value (P.V.) goes beyond the upper limit (HIGH
judgment), when the judgment goes beyond the lower limit (LOW judgment) and when the
judgment is within the acceptable range (GO judgment). Then, the judgment indicator and
judgment output are turned ON/OFF.
*1 When the output state of judgment output is Normally Open (default value) ON/OFF is
reversed for Normal Close.
*2 The judgment indicator ON/OFF condition can be changed.
"18. Display Color" (page 4-41)
*3
"Error Displays and Corrective Actions" (page A-4)
The tolerance setting value can be set either manually or automatically.
judgment
judgment output
*1
judgment indicator
*2
HIGH
GO
LOW
HI
GO
LO
HIGH
ON
OFF
OFF
Red
OFF
OFF
GO
OFF
ON
OFF
OFF
Green
OFF
LOW
OFF
OFF
ON
OFF
OFF
Red
Error
*3
ON
OFF
ON
Red
OFF
Red
DIN rail mount type (IL-1000/IL-1050)
Panel mount type (IL-1500/IL-1550)
Item
Setting Method
Manual setting
Directly enter the tolerance setting value (HIGH side
setting value, LOW side setting value).
Automatic
setting
(When other
than step count
filter)
Tolerance tuning
Detect the master workpiece and set the tolerance.
2 point tuning
Detect the good target and defective target and set the
tolerance.
Automatic
setting
(When step
count filter)
2 point tuning
Detect the step, set the tolerance.
1 point tuning
Detect the present P.V. (Judgment Value), set the
tolerance.
Point
When setting the tolerance setting value manually using the two-point
tuning, make sure to set "HIGH side setting value > LOW side setting value".
Judgment indicator