4-4 Calibration Function

4-51

IL-E

4

Setting V

ar

iou

s F
unct

ions

Main screen

1

Press [MODE] button and X button for approx. 2
seconds on the main screen at the same time.

[CALib] is displayed on the main display (upper level).

2

Press S / T button to set [CALc] and press X button.

Press [MODE] +
X

button

2 seconds or more
at the same time

3

Press S / T button to set [USr.3p] and press X button.

[SET] button

[

SEt1]

[SET] button

4

Workpiece A is set closer to the head connected to the
main unit allowing for the possible workpiece positional
deviation. Then set the "CALC value you wish to display
(1)" for that workpiece A by
S / T buttons.

The current internal measurement value is displayed on the main
display (upper level). [SEt1] and the "CALC you wish to display
(1)" are alternately displayed on the sub display (lower level).
IL-S025/IL-030/IL-S065/IL-065/IL-100

IL-300/IL-600

IL-2000

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

HOLD CALC CHECK

Setting value

Description

HEAd

The internal measurement value (R.V.) calibration is set.

CALc

The calculated value (CALC) calibration is set.

%#.E

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

%#.KD

HOLD

CALC

CHECK

Calculation calibration setting

7542

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

%%

HOLD CALC CHECK

Setting value

Description

Default value

dEFLt

Standard

c

USr.2p

User setting (Two-point calibration)

USr.3p

User setting (Three-point calibration)

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

HOLD

CALC

CHECK

CALC value

Setting you wish to display

Head connected

to main unit

Main unit side

Workpiece A

Head connected to
expansion unit

Item

Setting range

Default value

set1

-99.999 to 99.999

5.000

Item

Setting range

Default value

set1

-999.99 to 999.99

50.00

Item

Setting range

Default value

set1

-9999.9 to 9999.9

500.0