4 calibration function – KEYENCE IL Series User Manual
Page 102
4-4 Calibration Function
4-48
4
Setting V
ar
iou
s F
unct
ions
IL-E
When there is a difference between the calculated value (CALC value) and the actual
dimension of the object, the value can be corrected.
Both the zero point (offset) and tilt (span) are corrected.
Setting method (Two-point calibration of calculated value (CALC value))
Point
• If the following conditions are not satisfied, the screen below is
displayed when the setting is complete. Make the setting again.
•
• Workpiece is not within measurement range.
• The setting was changed during the calibration.
Main screen
1
Press [MODE] button and X button for approx. 2
seconds on the main screen at the same time.
[CALib] is displayed on the main display (upper level).
2
Press S / T button to set [CALib] and press X button.
Press [MODE] +
X
button
2 seconds or more
at the same time
3
Press S / T button to set [USr.2p] and press X button.
Calibration
Offset value A
CALC you wish to display (2) [
5'V]
CALC before calibration (2)
CALC you wish to display (1) [
5'V]
CALC before calibration (1)
Measurement value
Displayed value
'TT
LASER
BANK
0
1
2
3
HI
GO
LO
R.V. ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
5'6
HOLD CALC CHECK
0.5 ≤ (CALC you wish to display (2) ) - (CALC you wish to display (1))
(CALC before calibration (2)) - (CALC before calibration (1))
≤ 2
LASER
BANK
0
1
2
3
HI
GO
LO
R.V. ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC CHECK
Setting value
Description
HEAd
The internal measurement value (R.V.) calibration is set.
CALc
The calculated value (CALC) calibration is set.
%#.E
LASER
BANK
0
1
2
3
HI
GO
LO
R.V. ANALOG
HI
SHIFT
ZERO SHIFT
TIMING
LO
%#.KD
HOLD
CALC
CHECK
Calculation calibration setting
Setting value
Description
Default value
dEFLt
Standard
c
USr.2p
User setting (Two-point calibration)
USr.3p
User setting (Three-point calibration)