4-4 Calibration Function

4-48

4

Setting V

ar

iou

s F
unct

ions

IL-E

When there is a difference between the calculated value (CALC value) and the actual
dimension of the object, the value can be corrected.
Both the zero point (offset) and tilt (span) are corrected.

Setting method (Two-point calibration of calculated value (CALC value))

Point

• If the following conditions are not satisfied, the screen below is

displayed when the setting is complete. Make the setting again.

• Workpiece is not within measurement range.
• The setting was changed during the calibration.

Main screen

1

Press [MODE] button and X button for approx. 2
seconds on the main screen at the same time.

[CALib] is displayed on the main display (upper level).

2

Press S / T button to set [CALib] and press X button.

Press [MODE] +
X

button

2 seconds or more
at the same time

3

Press S / T button to set [USr.2p] and press X button.

Calibration

Offset value A

CALC you wish to display (2) [

5'V]

CALC before calibration (2)

CALC you wish to display (1) [

5'V]

CALC before calibration (1)

Measurement value

Displayed value

'TT

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

5'6

HOLD CALC CHECK

0.5 (CALC you wish to display (2) ) - (CALC you wish to display (1))

(CALC before calibration (2)) - (CALC before calibration (1))

2

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

HOLD CALC CHECK

Setting value

Description

HEAd

The internal measurement value (R.V.) calibration is set.

CALc

The calculated value (CALC) calibration is set.

%#.E

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

%#.KD

HOLD

CALC

CHECK

Calculation calibration setting

Setting value

Description

Default value

dEFLt

Standard

c

USr.2p

User setting (Two-point calibration)

USr.3p

User setting (Three-point calibration)