beautypg.com

KEYENCE IL Series User Manual

Page 47

background image

3-4 Setting the Tolerance Setting Value

3-15

IL-E

3

Basic Oper

ations

5

Measure the good target again and press the [SET] button. (LOW side 1st
point confirmation operation)

The internal measurement value (R.V.) is imported as a good target measurement
value.
[Loset] is displayed on the main display (upper level).

6

Measure the LOW side defective target and press the [SET] button. (LOW
side 2nd point confirmation operation)

The internal measurement value (R.V.) is imported as a measurement value for LOW
side defective target.
After [set] blinks on the main display (upper level), the judgment value (P.V.) is
displayed.
On the sub display (lower level), the median value of the good target value (1st point)
imported on step 5 and LOW defective target value (2nd point) is displayed.
Setting the LOW side setting value (lower limit) is complete.
The two-point tuning is complete.

Setting value for the HIGH side defective target

Setting value for the LOW side defective target

Setting value for the good target

Measurement upper limit value

Measurement lower limit value

LOW side setting value (lower limit)

HIGH side setting value (upper limit)