4-2 Basic Settings and Advanced Settings

4-9

IL-E

4

Setting V

ar

iou

s F
unct

ions

The averaging rate, the step count filter, and the high-pass filter data acquired by the
sampling cycle will be set.

„

Averaging rate

The average value is the moving average. If the measured values fluctuate, stable
measurements can be obtained by increasing the average count. Also, when selecting
[diff], the step function, or [HPF] the High-pass filter function will be enabled.

Point

• When the mutual-interference prevention function is enabled, the

sampling rate of the expansion unit is set equal to that of the main

unit, so the sampling rate setting screen for the expansion unit is
skipped.

• When the calculation function is enabled, the sampling rate of

expansion unit 1 is set equal to that of the main unit, so that the
sampling rate setting screen for expansion unit 1 is skipped.

Expansion unit 2 and following, units are each set individually.

• If the data between the heads is not synchronized when using the

calculation function or when the DL series is connected, disable

the mutual-interference prevention function and set the main unit's

sampling rate to a rate other than 0.33 ms. When the mutual-inter-
ference prevention function is enabled, or 0.33 ms is set, the tim-

ing for updating the main unit and expansion unit's internal data

will not match. In this case, when measuring a moving target using
the calculation function, the measured value may greatly differ

from the actual value.

• When the mutual-interference prevention function is enabled, 2 ms

or 5 ms cannot be set for IL-S025/IL-030/IL-S065/IL-065/IL-100, and
dEFLt

, 2 ms or 5 ms cannot be set for IL-300/IL-600/IL-2000.

• If the [FrEq] settings of the “3.AVE” of the [HPF] are set to 50 Hz/

100 Hz, the IL-S025/IL-030/IL-S065/IL-065/IL-100 cannot be set to

2 ms/5 ms, and the IL-300/IL-600/IL-2000 cannot be set to dEFLt/2
ms/5 ms.

• When “3.AVE” is [diFF], 5 ms cannot be set.

• When using in an environment that greatly interferes with the

beam, set 2 ms or 5 ms.

3. Averaging rate, Step count filter, High-pass filter

#8'

Item

Setting range (rate)

Default value

Averaging rate

1

/ 2 / 4 / 8 /16 / 32 / 64 /128 / 256 / 512 /1024 / 2048 / 4096 /

diFF

/ HPF

16

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

#X'

GO

HOLD

CALC

CHECK

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

#X'

GO

HOLD

CALC

CHECK

F'(.V

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

#./

GO

HOLD

CALC

CHECK

3. Averaging rate,
Step count filter,
High-pass filter

Averaging
setting screen

S

/ T

button

[MODE]
or X

(page 4-9)

(page 4-9)

4. Alarm setting

(page 4-13)