4-2 Basic Settings and Advanced Settings

4-19

IL-E

4

Setting V

ar

iou

s F
unct

ions

*1

"11. External Input" (page 4-30)

*2 The main unit (IL-1000/IL-1500) displays the calculation value (CALC value) when the

calculation function is used.

“4-3 Calculation Function” (page 4-42)

*3 If the R.V. value (or CALC value) causes an alarm even once during the sampling period, the

hold results will be "-----".

Peak-to-peak hold

The difference between the maximum value and minimum value of the specified period
(sampling period) is held as a judgment value (P.V.).

„

When "Level" is set for "7. Timing Input" (page 4-23)

The sampling period is from the falling edge of the timing input to the next rising edge. On
the rising edge, the difference between the maximum value and minimum value of the
internal measurement value (R.V.) during the sampling period is held as a judgment value
(P.V.). When the timing input is turned OFF, the next sampling period begins.

*1

"11. External Input" (page 4-30)

*2 The main unit (IL-1000/IL-1500) displays the calculation value (CALC value) when the

calculation function is used.

“4-3 Calculation Function” (page 4-42)

*3 If the R.V. value (or CALC value) causes an alarm even once during the sampling period, the

hold results will be "-----".

Display value

Time

Timing Input

Sampling period

Sampling period

ON

OFF

Power ON or
Reset input *

1

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

p : Maximum

b : Minimum

b

b’

p’

(p’-b’)

p

(p-b)

Judgment value (P.V.)
Internal measurement value (R.V.)*

2