5. Methods and determinations

796 Titroprocessor, Instructions for use

91

91

Statistics and sample series

In certain cases it may be a good idea to set the statistics limit to a lower
value than the total number of samples. If double or triple determinations
are to be carried out on the same sample within a sample series and
statistical data are to be calculated for each of them then the statistics limit
should be set to 2 or 3 respectively.
The statistical calculations are carried out for each sample but with these
settings the statistics are reset after each 2 or 3 determinations and the
working memory cleared. In this way statistical calculations are obtained for
double or triple determinations.

Warning! Statistical results are only calculated according to formulas which
have been defined as a 'Statistics formula' (formula type 'stat.') in the
formula editor. The statistics should be initialized before the start of a
sample series.