8. Command reference

796 Titroprocessor, Instructions for use

189

189

PUMP

Pump control

Different pumps can be controlled depending on the sample changer ver-
sion. A rinsing and a aspiration pump can be installed on one changer
tower. These can be switched separately. The duration of rinsing and suc-
tion processes can be comfortably controlled with the line time of the

PUMP

command.
External pumps, e.g. for removing titration solutions containing solids, can
be controlled via remote output leads (remote socket, see

CTRL com-

mand).

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

PUMP

Address 1.rinse State off

Address

Pump address

1.rinse | 1.aspir | 2.rinse | 2.aspir

The address of a pump consists of the specification of
the changer tower (1 or 2) and the function of the pump
(rinsing=pump 1, aspirate=pump 2). It is valid for the
currently active changer.

State

Pump status

on | off

A pump can only be switched on or off. The pumping
duration can be controlled with the line time of the PUMP
command. Example:

t/s Command … Parameters
----- --------- ---------------
5 PUMP … State on
PUMP … State off

STIR_C

Sample changer stirrer control

In contrast to the stirrer connections of the Titroprocessor
itself, the stirrer speed can be controlled on the 717 Sam-
ple changer.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

STIR_C

Address 1 Stirring rate 15

Address

Stirrer address

1…4

The stirrer address always refers to the active changer.
With 1-tower changer models only 2 stirrers (722 Rod
stirrer or 741 Magnetic stirrer) can be connected.