3. Configuration
796 Titroprocessor, Instructions for use
45
45
3.3.1 Setting the RS232 interfaces
Baud rate
Transmission speed
Data bit
Number of data bits
Stop bit
Number of stop bits
Parity
Type of parity check
Handshake
Type of protocol
Receive mode
Allows or blocks the reception of transmitted data
3.3.2 Balances and printers
Enter the type of balance or printer used in the column allocated to the
RS232 interface to which the corresponding instrument is connected, see
Titroprocessor rear panel.
Balance
Manufacturer or type of connected balance
Printer
Manufacturer or type of printer
– Select 'HP-Deskjet' for inkjet printers (with
RS232 interface) from Hewlett-Packard or HP-
compatible laser printers
– Select 'Epson 80*60' or 'Epson 60*72' for
Epson printers or matrix printers from other
manufacturers. (80*60 means a graphics resolution
80 dpi by 60 dpi).
– Select 'IBM Proprinter' if your printer does
not produce a satisfactory printout with the above
settings.
If the Titroprocessor 796 does not have a built-in thermal printer then a
connection socket 'Printer' will be found on the rear panel of the instrument.
A normal commercial printer can be connected to this socket with a parallel
Centronics interface. This printer should be defined under 'Printer
port'
.
³ communication Int. 1 ³
³ commun. mode LIMS ³
³ ³
³ Printer port ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ Printer ---- ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ
3.3.3 Data communication settings
The Titroprocessor 796 can communicate with a PC or a LIMS system via
the RS232 interfaces. This ensures that reports or determination files, etc.
can be stored on a computer. It is also possible to operate the
Titroprocessor in a key-simulation mode with the aid of a personal computer
and suitable software.