Metrohm 796 Titroprocessor User Manual
Page 259
12
12. . Troubleshooting
Troubleshooting
796 Titroprocessor, Instructions for use
253
253
SCAN+CTRL
SCAN+CTRL 1:
Remote interface X not present
The selected 'Remote' interface has not been installed or is not able to func-
tion.
STATIST
STATIST 9:
determ. running, reset not possible
The statistics cannot be reset while a determination is being carried out, as
the working memory would be completely cleared. This would also affect
the determination being carried out.
STORAGE
STORAGE 1:
file is locked
The file mentioned is protected against deletion or overwriting. The protec-
tion can be removed in the file manager with the function [Un/Lock file].
STORAGE 2:
file not found
A file with the file name mentioned cannot be found in the selected memory
area. Check whether the file name has been entered correctly.
STORAGE 5:
inadmissible destination for access control files
Access control files cannot be loaded into the working memory directly.
They are loaded under [Config.] [Access control] [File].
STORAGE 6:
Inadmissible destination for configuration files
Configuration files cannot be loaded into the working memory directly. They
are loaded under [Config.] [File].
STORAGE 7:
Inadmissible destination for rack files
Rack files cannot be loaded into the working memory directly. They are
loaded under [Config.] [Sample changer] [Rack File].
STORAGE 8:
invalid determination
The determination file is damaged and can no longer be read. Delete it.
STORAGE 9:
switch on the data "Receive mode"
For the required action the data receive mode of the RS232 interface must
be switched on. This can be carried out under [Config.] [Interfaces].