Metrohm 796 Titroprocessor User Manual
Page 33
22. . Installation
Installation
796 Titroprocessor, Instructions for use
27
27
2.6.6 Connecting a balance
RS 232 Interface 1
RS 232 Interface 2
G
The following balances can be connected to an RS 232
interface of the 796 Titroprocessor:
Balance
Cable
Sartorius MP8, MC1
6.2134.060
Mettler AB, AG, PR
LC-RS25, supplied with balance
Mettler AM, PM
from Mettler: ME 33995: green wire to pin 2, brown to pin
3, white to pin 7, yellow to pin 20 of the 25-pole plug
+ 6.2125.010 25-pole/9-pole adapter
Mettler interface 016
cable supplied with interface 016: red wire to pin 3, white
wire to pin 7 of the 25-pole plug
+ 6.2125.010 25-pole/9-pole adapter
Mettler interface 011 or 012
6.2125.020 + 6.2125.010
Mettler AT
from Mettler: ME 33995: green wire to pin 2, brown to pin
3, white to pin 7, yellow to pin 20 of the 25-pole plug
+ 6.2125.010 25-pole/9-pole adapter
AND types ER-60, 120, 180, 182
types FR-200, 300
types FX-200, 300, 320
with RS 232 interface (OP-03)
6.2125.020 + 6.2125.010
Precisa, balances with RS 232C in-
terface
6.2125.080 + 6.2125.010
The type of balance must be entered in the Titroprocessor on the configura-
tion page under [Interfaces].