8.10 Sample changer commands

796 Titroprocessor, Instructions for use

186

186

On beaker error

Behavior with missing sample beaker

MOVE_S | display

If a beaker sensor finds that a sample beaker is missing
then this instruction can be used to carry out two sub-
sequent actions automatically:

display

An error message is displayed and
the operator must take action. The
instrument moves to the

'hold' condit-

ion and the changer can be operated
manually.

MOVE_S

The previous

MOVE command is

carried out again; the changer "looks
for" the next sample beaker. A pre-
vious

SAMPLE command will be

taken into account, i.e. with
'SAMPLE +2' the next but one sample
beaker will be moved to.

SAMPLE

Alteration of the run variable SAMPLE

The run variable SAMPLE is used as a sample counter during the process-
ing of a series of samples. It always refers to the current sample position in
the sample rack. It can be set manually with the

'Sample pos.' entry on the

Titroprocessor main page. However, a

SAMPLE command will overwrite

this entry. When a sample has been processed the run variable will be
automatically increased by

1 so that in a method the next sample beaker

can always be brought into position with the command

MOVE_S.

This run variable, and therefore the particular valid sample position, can be
specifically altered with this

SAMPLE command. If the SAMPLE command

is used in a method then the automatic

SAMPLE variable increase is

switched off.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

SAMPLE

Function + Value 1

Function

Type of SAMPLE variable alteration

+ | – | =

The SAMPLE variable can be increased (

'

+

') or de-

creased (

'–'). Allocation of a specific value ('=') (setting

a particular sample position) only makes sense in an
(

OMOVE) start sequence.