Metrohm 796 Titroprocessor User Manual
Page 37
22. . Installation
Installation
796 Titroprocessor, Instructions for use
31
31
2.6.9 Devices on the remote interface
External peripheral devices such as pH meters, pumps, etc. can be con-
trolled via the remote interface (25-pole socket) of the Titroprocessor. Each
717 Sample changer also has a remote socket which can be controlled via
the
'External bus'.
Remote
E
8 lines are available for signal output (Output 0-7). The remote socket of the
717 Sample changer has 14 output lines available (Output 0-13).
For receiving signals (e.g. the "ready" signal from a Metrohm instrument)
there are 8 lines available (Input 0-7).
The pin occupancy of the remote socket:
Output 6
Output 7
Output 4
Output 2
Output 0
Output 8
Output 9
Output 10
Input 1
Input 3
Input 5
Input 7
Output 11
0 Vo lt
+5 Volt
Output 5
Output 3
Output 1
Output 12
Output 13
Input 0
Input 2
Input 4
Input 6
0 Vo lt
The control lines of the 796 Titroprocessor are purely signal leads (TTL-level)
and should not be misused for supplying devices.
Remote connections can be used with specially assembled cables to solve
complex connection problems in automated systems.
Connection of Metrohm instruments is solved in the following manner with
6.2141.020 Standard remote cable
796
Metrohm instrument
³
796
Metrohm instrument
³
Output 0
–––––––
Input 0
³
Input 0
–––––––
Output 0
Output 1
–––––––
Input 1
³
Input 1
–––––––
Output 1
Output 2
–––––––
Input 2
³
Input 2
–––––––
Output 2
Output 3
–––––––
Input 3
³
Input 3
–––––––
Output 3
Output 4
–––––––
Input 4
³
Input 4
–––––––
Output 4
Output 5
–––––––
Input 5
³
Input 5
–––––––
Output 5
Output 6
–––––––
Input 6
³
Input 6
–––––––
Output 6
Output 7
–––––––
Input 7
³
Input 7
–––––––
Output 7
Output 8
–––––––
Pin 6
Output 9
–––––––
Pin 7
Output 10 –––––––
Pin 8
Output 11
–––––––
Pin 13
Output 12
–––––––
Pin 19
Output 13
–––––––
Pin 20
Output lines 8…13 can only be used
with the 717 Sample changer, but are
not used as input lines for Metrohm in-
struments. They are laid 1:1 on Pin
6…8,13,19…20.
+5V
active= low
inactive = high
active = low
inactive = high
Inputs