13.3 Key cards and access control
796 Titroprocessor, Instructions for use
272
272
13.3 Key cards and access control
The user dialog of the Titroprocessor 796 can be limited according to the
wishes of the operator. For example, in order to avoid accidental deletion or
alterations of methods by unauthorized persons it is possible to ‘hide’ or
block softkeys and several keyboard keys. This results in a simplified user
dialog. The configuration of the user dialog can be stored in a file on a
memory card. In addition to the modified user dialog, certain status settings
(user name, statistics on/off, source location of methods etc.) are stored in
this 'access control' file. Memory cards with an access control file can be
used as key cards. The 796 Titroprocessor may be configured, so that ac-
cessing the instruments functions is only possible by use of a key card.
Every user may have his own dialog and basic settings when using a per-
sonally configured key card.
The main page of the Titroprocessor could appear similar to that below:
This means that no methods or determination results can be altered.
The status settings can be individually predefined.
ЪДДД
Status
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ Run number
000
Determination name ³
³ ³
³ Sample silo on Next sample in silo line 1 ³
³ Sample entry in silo line 1 ³
³ ³
³ Auto start off ³
³ ³
³ Statistics off ³
³ ³
³ Save determ. data off ³
³ ³
³ ³
³ Sample changer ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ
1999-12-21 10:17:52 MAIN PAGE 796 Titroprocessor * ready *
Method 3455.mth Chloride in Tap Water
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД
ЪДДДДДДДДДД Application note ДДДДДДДДДДДДДДД¿
User ³ Chloride in Tap Water ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД³
³ ³
³ Pipet 100.00 mL tap water exactly in a ³
³ 150 mL beaker. ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДЩ
Ident1 Ident2 Ident3 Size Unit
Sample
Remark
ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿
³ ³ ³ ³ ³ ³Show ³ ³ ³ ³Card ³ ³
³ ³ ³ ³ ³ ³curves ³ ³ ³ ³Sign on³ ³
АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ