7. Titration and measuring modes
796 Titroprocessor, Instructions for use
121
121
stat.
The result enters the statistics.
live
The result is shown in the trace win-
dow during the determination and
recalculated each time a new entry
is made in the measuring point list.
Please note that in titrations the live
value may not be quite as up-to-
date as the actual value.
Softkeys
[Copy line]
copies the contents of the selected line into a local
buffer memory.
[Cut line]
copies the contents of the selected line into a local
buffer memory and deletes the selected line.
[Paste line]
inserts a new line before the selected line and enters
the contents of the local buffer memory in it.
[Insert line]
inserts a new empty line before the selected line.
[Delete line]
deletes the selected line.
7.10.1 System variables
The following system variables can be used in formulas.
(xx=any 2 numbers or letters, #=any number)
Rxx
Result
RT1
Result for validation
RT2
Theoretical result for validation
RE
Result for calculating the system error in validation
Xxx
Intermediate result
C##
Common variable (of the actual titration or measurement)
EP#
Endpoint volume in mL
EM#
Endpoint measurement (unit depends on mode)
EF#
Endpoint flag; endpoint found yes/no (1/0)
F#
Fixed endpoint volume in mL
FM#
Fixed endpoint measurement (unit depends on mode)
FF#
Fixed endpoint flag; fixed endpoint reached yes/no (1/0)
MNRxx
Mean value of results Rxx in statistics memory
MNXxx
Mean value of intermediate results Xxx in statistics memory
MNC##
Mean value of common variables C##
ID1
Sample identification 1
ID2
Sample identification 2
ID3
Sample identification 3
SS
Sample size (value without units)
DD
Determination duration in s, time for whole determination
MD
Measuring duration in s; time for actual titration or measure-
ment
SD
Starting duration in s; length of start phase of a mode
IT
Initial temperature in °C
ST
Starting temperature in °C (after start phase), only for auto-
matic measurements.
CT
Closing temperature in °C, only for automatic measurements