3 mains connection – Metrohm 796 Titroprocessor User Manual
Page 23
22. . Installation
Installation
796 Titroprocessor, Instructions for use
17
17
WARNING - Fire Hazard -
with the same type and rating of fuse
For continued protection replace only
50-60Hz 160VA
2A(TH)
115V
230V
230V
115V: 100V...120V 10%
230V: 220V...240V 10%
2.3 Mains connection
2.3.1 Fuses
Two fuses
S1
S1
and
S2
S2
are built into the 796 Titroprocessor as standard;
both are type 2 ATH (2 A, slow-blow, with high switching capacity, Metrohm
order no. U.600.0107).
Make sure that the instrument is never operated with any different type
of fuse as otherwise there is a fire hazard!
To change blown fuses proceed as follows:
Exchange fuse S1 (zero conductor, 2 ATH)
•
Remove mains cable from mains connector.
•
Use a screwdriver to turn fuse holder
S1
S1
to the left until it can be pulled
out.
•
Pull out the fuse holder, remove the blown fuse and replace it with a
spare fuse (2 ATH).
•
Insert the fuse holder in the instrument, press it down with a screwdriver
and then turn it to the right to fix it in position.
Exchange fuse S2 (phase, 2 ATH)
•
Remove mains cable from mains connector.
•
Use a screwdriver to push fuse holder
S2
S2
upwards until it can be pulled
out.
•
Pull out the fuse holder, remove the blown fuse and replace it with a
spare fuse (2 ATH).
•
Insert the fuse holder in the instrument and push until it clicks into place.
Fuse holder
Mains voltage selector
Earth socket
Mains switch
Rear view
S1
S2
Mains connection