Metrohm 796 Titroprocessor User Manual
Page 83
5. Methods and determinations
796 Titroprocessor, Instructions for use
77
77
Report printout
yes | no
With 'yes' a report will be printed out on all printers defined
in the configuration. This could be both the internal printer
as well as an external printer connected to the 'Printer'
connection or an RS232 interface.
send
Report output via an RS232 interface
yes | no
A report can be transferred directly to a personal
computer via an RS232 interface. The particular RS232
interface to which the receiving instrument is connected
must have been defined as being the communication
interface in the interface configuration and the data
transfer mode 'LIMS' must be switched on. Take care that
the transmission parameters (baud rate, etc.) of the
receiving instrument are set to the same values as on the
Titroprocessor 796.
FF
Page feed (form feed) at printer
yes | no
For formatting the printout of several report blocks it is
possible to provide each individual report with a following
page feed. A 'Form Feed' only affects the printer. Please
note that the internal thermal printer will only carry out a
form feed if 'Enable auto FF' is switched on in the
configuration of the interfaces.
Print signature line
Operator name and signature line
yes | no
A signature line with the operator’s name can be printed
out at the end of the complete report.
Final form feed
Page feed after last report line
yes | no
If reports are to be printed out on a page-oriented external
printer (inkjet printer, laser printer) then a form feed must
be carried out after the last report in order for the final
page to be ejected.