Metrohm 796 Titroprocessor User Manual
Page 126
7.10 Calculations / Formulas
796 Titroprocessor, Instructions for use
120
120
Remark
Formula name
11 characters
Each formula can be given any name. The name only has
a descriptive character.
Formula
Calculation formula
35 characters
A formula consists of a variable (final result, intermediate
result or common variable), which is allocated to the re-
sult of the calculation and the calculation formula itself,
which can consist of mathematical operators, numerical
values, common variables and system variables. Results
of previous calculations also belong to the system vari-
ables.
Calculation results can be allocated the following vari-
ables:
R00…RZZ
Final result, is given on the result page
and in reports; can be used for calcula-
tions, statistics and live displays.
X00…XZZ
Intermediate result, is not shown; can
be used for calculations, statistics and
live displays.
C01…C19
Common variables, are given on the
result page and in reports; can be used
for calculations, statistics and live dis-
plays. They are globally available for
subsequent determinations.
Unit
Unit of the result
% | ppm | g/L | mg/mL | mg/pc |
mol/L | g | mg | mL | s | mL/min |
max. 6 characters
Any other unit can be entered apart from the selection of
units given above,.
Dec
Decimal place in the result
0…5
The result of the calculation will be rounded off to the
number of decimal places given above and this rounded-
off value enters the statistics. If the result is used internally
for other calculations then the original and completely ac-
curate value will be used.
Type
Type of result
normal | stat. | live
The type of result determines how it will be shown:
normal
The result is shown in the result
window and in the report.