12.1 Error messages
796 Titroprocessor, Instructions for use
248
248
METHOD 8:
incomplete (OMOVE sequence, see line XX
The start sequence (OMOVE of the method has not been completely de-
fined. A sequence must always be concluded with the corresponding com-
mand, in this case OMOVE).
METHOD 9:
incomplete (CMOVE sequence, see line XX
The closing sequence (CMOVE of the method has not been completely de-
fined. A sequence must always be concluded with the corresponding com-
mand, in this case CMOVE).
METHOD 10:
incomplete (CAL sequence, see line XX
The calibration sequence (CAL of the current method has not been com-
pletely defined. A sequence must always be concluded with the corre-
sponding command, in this case CAL).
METHOD 11:
incomplete (CASE sequence, see line XX
METHOD 12:
incomplete (CASE sequence, see line XX
The condition sequence (CASE of the current method has not been com-
pletely defined. A sequence must always be concluded with the corre-
sponding command, in this case CASE).
METHOD 13:
incorrect OMOVE / CMOVE structure, see line XX
Non-permissible nests of command sequences have been programmed in
the method. A sequence must always be concluded with the corresponding
command before another sequence can be started.
METHOD 14:
line XX: illegal nesting, check method!
Non-permissible nests of command sequences have been programmed in
the method. A sequence must always be concluded with the corresponding
command before another sequence can be started.
MONITORING
MONITORING 1:
no more samples in silo
There are no further samples in the silo. The silo pointer is located at the
bottom line. Set the '<-- Out pointer' again. Silo 'rst' on the main page or in
the
MONITORING 2:
missing method XXXXXXXX
A method given in the silo cannot be loaded because it is stored neither in
the internal memory nor on the data card.
MONITORING 3:
check method XXXXXXXX
The given method is faulty and cannot be carried out.