6 kft – Metrohm 796 Titroprocessor User Manual
Page 165
8. Command reference
796 Titroprocessor, Instructions for use
159
159
8.6 KFT
Two different
KFT modes are available for titrimetric water determination
with Karl Fischer reagents; they differ only in the measuring method:
KFT_UPOL*
measured quantity µA (selectable pol.voltage in mV)
KFT_IPOL*
measured quantity mV (selectable pol.current in µA)
The following information applies to both
KFT modes. Any variations are
noted.
KFT_IPOL*
Karl Fischer titration
Reagent Sensor
Concentr. Meas. input Apol
Dos. drive A1
applies to
⇒
KFT_IPOL*, KFT_UPOL*
KFT modes produce the same system variables as DET modes, see page
129. In addition the following variables are provided for calculating a drift
volume reduction (see page 155):
DOD
Dosing duration during the titration, cumulative
ITD
Time duration between conditioning and titration
Assignment parameters for KFT modes:
The assignment parameters (Comment, Reagent, etc.) for the
KFT modes
are the same as those for the
DET modes, see page 129f.
8.6.1 Specific parameters for KFT modes [ * Param ]
The specific parameters for the respective
KFT modes can be found under
the softkey [* Param ].
ЪДДД
KFT_IPOL* Parameters
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ Start conditions ³
³ Volume off Pause 1 0 s ³
³ Pause 2 0 s ³
³ Extract.time 0 s ³
³ ³
³ Titration conditions ³
³ Direction neg. Temperature 25.0 øC ³
³ End-point 250 mV Time recording on ³
³ Adjustment medium Derivation off ³
³ Meas.interval off s ³
³ I(pol) 50 µA ³
³ Stop conditions ³
³ Volume absolute Stop time off s ³
³ Size 99.99 mL Mpt. overflow off ³
³ Fill. rate max mL/min ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ