Metrohm 796 Titroprocessor User Manual
Page 138
8.2 DET
796 Titroprocessor, Instructions for use
132
132
The various conditions governing the discontinuation of initial dosing can be
combined at will.
N.B.: Initial dosing is discontinued on attainment of any of the start condi-
tions, be it the defined absolute/relative start volume, the defined measured
value or the signal slope. The actual titration will then start immediately after
the pre-set pause.
predef. start volume
meas. value
(reached)
Slope
Vol [mL]
U [mV]
actual initial dosing
The volume actually dosed is filed in the
SV system variable and can be
called up for calculations.
Volume
Definition of the start volume
absolute | relative | off
The defined start volume can be entered as an absolute
volume, or as a relative volume as a function of the sample
size (recommended). Various other fields* will need to be
edited depending on what is entered in this field
(see be-
low).
Size
*
Absolute volume (with abs. start volume)
0…999.99 mL
If an absolute start volume is entered, the start volume
should be kept small enough to ensure that sufficient vol-
ume remains up to the equivalence point for the dynami-
cally controlled titration.
Factor
*
Calculation factor (with rel. start volume)
–99999…0…99999
The actual volume of the start volume to be added will be
calculated as follows:
Volume in mL = sample size
∗∗ factor
.
The maximum start volume that can be added is 1000 mL.
If the calculation results in a higher value, the start volume
is automatically reduced to 1000 mL.