8.3 MET
796 Titroprocessor, Instructions for use
144
144
8.3 MET
There is a choice of four different
MET modes for controlled titrations with
fixed volume steps. They differ only with regard to the measuring quantity:
MET_PH*
Measuring quantity pH
MET_U*
Measuring quantity mV
MET_UPOL*
Measuring quantity µA (adjustable pol. voltage in mV)
MET_IPOL*
Measuring quantity mV (adjustable pol. current in µA)
The following information applies to all
MET modes. Discrepancies are indi-
cated.
MET_PH*
monotonic equivalence point titration
Reagent Sensor
Concentr. Meas. input A1
Dos. drive A1
applies for
⇒
MET_PH*
, MET_U*, MET_UPOL*, MET_IPOL*
MET modes create the same system variables as DET modes, see
page 129.
Assignment parameters of the MET modes:
The assignment parameters (comment, reagent … etc.) of the
MET modes
coincide with those of the
DET modes, see page 130f.
8.3.1 [ * Param ] Specific parameters for MET modes
The specific parameters of the corresponding
MET mode will be found un-
der the [* Param ]softkey.
ЪДДД
MET_PH* Parameters
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ Start conditions ³
³ Volume off Meas.val. pH off ³
³ Slope off pH/mL ³
³ Dos. rate max mL/min Pause 0 s ³
³ ³
³ Titration conditions ³
³ Adjustment medium Time recording on ³
³ Temperature 25.0 øC ³
³ ³
³ Stop conditions ³
³ Volume absolute Meas.val. pH off ³
³ Size 99.99 mL EP# 9 ³
³ fill. rate max mL/min Stop time off s ³
³ Mpt. overflow on ³
³ ³