Metrohm 796 Titroprocessor User Manual
Page 297
13. Appendix
796 Titroprocessor, Instructions for use
291
291
Ambient temperature Nominal operating range
5…40 °C
Storage, transport
–25…60 °C
Ambient atmospheric at
≤40 °C
max. 95% relative humidity
humidity
at 45°C
max. 70% relative humidity
at 60°C
max. 35% relative humidity
Titration and measurement modes
DET
Dynamic equivalence-point titration; U, pH, I
pol
, U
pol
MET
Monotonic equivalence-point titration; U, pH, I
pol
, U
pol
SET
Titration to preset endpoint; U, pH, I
pol
, U
pol
SEC
Conditioning to preset endpoint; U, pH, I
pol
, U
pol
KFT
Karl Fischer endpoint titration; I
pol
, U
pol
KFC
Conditioning according to Karl Fischer; I
pol
, U
pol
MEAS
Measurement; U, pH, I
pol
, U
pol
, T
CAL
pH calibration
Memories
Program memory
EPROM, max. 4 MB
Data memory
2 MB SRAM, battery backed
Determination memory for up to 20 determination data records each with
up to 5 titrations or measurements
Method memory
for freely definable sequence, up to 99 command
lines
Silo memory
sample data memory for up to 160 sample data
records
File memory
free RAM area for the storage of determinations,
methods, sample data tables
Memory card
Type
Memory card according to standard JEIDA 4.X /
PCMCIA 2.X (68 pins)
Memory capacity
2 MB (6.2245.030),128 kB (6.2245.010);
other versions up to 2 MB applicable
Safety specifications
Construction and testing in accordance with IEC 1010 / EN 61010 / UL
3101-1, protection class I
Degree of protection IP 52
The instruction manual contains information and warnings which the user
should follow to guarantee the safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference
The 796 Titroprocessor complies with the basic specifications EN 50081-1
01.92, EN 55011 (class B), EN 55022 (class B), and NAMUR.
Noise immunity
The standards EN 50082-2: 1995, IEC801-2 up to IEC801-6, EN 60555-2
and NAMUR are complied with.