Metrohm 796 Titroprocessor User Manual
Page 181
8. Command reference
796 Titroprocessor, Instructions for use
175
175
MEAS_T*
Temperature measurement
Sensor Equilibr.time 1 s
Meas. input AT
applies to
⇒
MEAS_T*
No drift control is available in the temperature measuring mode
MEAS_T. If
necessary, measurement acceptance can be delayed with the equilibrium
time in order for the sensor to achieve equilibrium temperature.
The other settings correspond to those of the other
MEAS modes.
Equilibr.time
Delay period before measurement acceptance
0.1…1…9999 s | auto | off
For
MEAS_T the whole equilibrium time is effective for
each measurement acceptance. The measured value is
accepted when the equilibrium time has elapsed.
With 'off' and 'auto' the equilibrium time is switched
off and the measurement is accepted immediately.
For multiple measurements the equilibrium time must
be shorter than the measuring interval as otherwise the
timing pattern for the measuring intervals cannot be fol-
lowed.
8.8.1 Standard measurement
A standard measurement is a single measurement in which a single meas-
ured value is recorded according to the measurement parameters and then
terminated. The measuring procedure (standard measurement / multiple
measurement) can be selected under the extended parameters (softkey
[*Param ]
) together with further data recording parameters.