3. Configuration
796 Titroprocessor, Instructions for use
37
37
Softkeys
[Device
header]
opens the dialog window for the entry of an
instrument-specific, four-line header, which will be
printed out before the report of a determination.
[File]
opens the file dialog window for storing and loading
configuration files. It is recommended that all the
instrument configuration settings are stored on a
data card in this way. In a configuration file all the
subordinate areas on the configuration page are
stored in addition to the common variables. The rack
tables of special racks are not stored and, if
necessary, must be separately loaded with the help
of rack files, see page 41.
In this way it is possible to reconstruct the whole
configuration of the Titroprocessor at a later date, or
to install the same settings on different instruments.
When loading a configuration file the individual
configuration areas can be selected for individual
take-over.
[Access
control]
opens, after the corresponding password has been
entered, the configuration page for definition of the
access control. Individual keys on the keyboard as
well as most of the softkeys on the dialog pages can
be blocked or blinded out. Details can be found on
page 272ff.
[Interfaces]
opens the configuration window of the interfaces.
The transfer parameters of the RS232 interfaces can
be entered here and the definition of balances and
printer set. Details can be found on page 44.
[Config
idents.]
opens the configuration window of the sample
identification fields. The titles of the fields (Ident1…3)
can be freely selected. The field types determine
whether the contents of the sample identification
fields are to be interpreted as text or as numerical
values. In the latter case the sample identifiers
(system variables
ID1…3) can be used in calculation
formula, see page 119ff.
If the sample silo is switched on then the definitions
of the sample identification fields in the sample silo
are valid. See also page 97.
[Buret
unit]
opens the configuration window of the buret units. In
order to use the dosing functions preparation
(
PREP) and empty (EMPTY) in an optimal way it is
necessary to define buret units in which the tubing