13.1 Diagnosis
796 Titroprocessor, Instructions for use
266
266
13.1.9 To check Analog Interface A
With this diagnostic step the analog inputs of Interface A can be checked for their function-
ality. For this test the testbox with article number 1.767.0010 is required.
1.
Prepare instrument for diagnostic test (see chap. 13.1.2).
2.
Ä
On the screen there appears a window with the choice of selecting between analog interface A or B.
3.
With the cursor key, mark the letter A to select analog interface A.
4.
5.
Connect Simulator to Titroprocessor accord. to Fig. 2.
6.
Ä
The measurements are continuously updated and presented as follows::
Test date: ??-??-?? ??:??:??
Analog input measuring test
!!!!!!!!!!!!!!!!!!!!!!!!!!!
!!!!!!!!!!!!!!!!!!!!!!!!!!!
Setting polarizing to: U=300 mV, I=10 uA
Display values as average of 6 meas. cycles
Analog interface A, Calib date: ??-??-?? ??:??:??
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
Input 1 : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C
Input 2 : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C
Input Diff : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C
Input Ipol : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C
Input Upol : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C
Fig. 1
7.
Check the following measuring values:
column C (see Fig. 1)
column E (see Fig. 1)
Input 1 :
+U / direct
±
0.4 mV
Pt100
±
0.3°C
Input 2 :
+U / 1G
±
0.4 mV
Input Diff :
0 mV
±
4.0 mV
8.
Connect Simulator to Titroprocessor accord. to Fig. 3.
column C (see Fig. 1)
Input 1 :
+U / 1G
±
0.4 mV
Input 2 :
+U / direct
±
0.4 mV
Input Diff :
0 mV
±
4.0 mV
DIAGNOSIS WINDOW
A
B
C
D
E