Appendix, Quit, All inp measure – Metrohm 796 Titroprocessor User Manual
Page 273: The test is quit by pressing the key twice
13. Appendix
796 Titroprocessor, Instructions for use
267
267
9.
Connect Simulator to Titroprocessor accord. to Fig. 4.
column C (see Fig. 1)
Input 1 :
-U% / direct
±
0.4 mV
10.
Connect Simulator to Titroprocessor accord. to Fig. 5.
column C (see Fig. 1)
Input 2 :
-U% / direct
±
0.4 mV
11.
Connect Simulator to Titroprocessor accord. to Fig. 6 and place the cover over the solar
cell.
column C (see Fig. 1)
Input Ipol :
143 mV
±
20 mV
Input Upol :
21 µA
±
3.0 µA
12.
13.
Connect Simulator to Titroprocessor accord. to Fig. 7.
14.
column E (see Fig. 1)
Input 1 :
Pt1000
±
0.3°C
15.
The test is quit by pressing the
1
Ind A2 Ind A1 Ref
Pol
Pt100/1000
Pt100
Pt1000
-U%/direkt +U/direkt +U/1G
726
Simulator
1
Ind A2 Ind A1 Ref
Pol
Pt100/1000
Pt100
Pt1000
-U%/direkt +U/direkt +U/1G
726
Simulator
1
Ind A2 Ind A1 Ref
Pol
Pt100/1000
Pt100
Pt1000
-U%/direkt +U/direkt +U/1G
726
Simulator
Fig.
Fig. 22
Fig.
Fig. 33
Fig.
Fig. 44
1
Ind A2 Ind A1 Ref
Pol
Pt100/1000
Pt100
Pt1000
-U%/direkt +U/direkt +U/1G
726
Simulator
1
Ind A2 Ind A1 Ref
Pol
Pt100/1000
Pt100
Pt1000
-U%/direkt +U/direkt +U/1G
726
Simulator
1
Ind A2 Ind A1 Ref
Pol
Pt100/1000
Pt100
Pt1000
-U%/direkt +U/direkt +U/1G
726
Simulator
Fig.
Fig. 55
Fig.
Fig. 66
Fig.
Fig. 77