12.1 Error messages
796 Titroprocessor, Instructions for use
252
252
For connection problems with external instruments please obtain advice
from the service department of your local Metrohm agency. Mention the
fault number and the error message text.
RS232 14:
interface X:Busy, remote keyb. mode not possible
The RS232 interface is currently occupied. Wait until the current task has
been completed and then alter the settings.
RS232 15:
interface X:unexpected error
A fault has occurred which cannot be specified more closely. Please con-
tact the service department of your local Metrohm agency.
RS232 16:
interface X, switch on the receive mode
The selected function requires the data receive mode of interface X. Switch
it on under [Config.] - [Interfaces].
RS232 17:
keySEND mode not possible with peripheral devices
The 'keySEND' mode is switched on and transfers all key actions via the
communication interface. This can affect the functioning of any printer, bal-
ance or other instrument connected to this interface. If the 'keySEND' func-
tion is required then a separate interface should be used for it.
RS232 18:
interface X, SCAN_RS not possible with keyCTRL mode
When the key simulation mode keyCTRL is switched on, no SCAN com-
mand can be carried out via the same interface. If the
tinue) is pressed, the procedure will be resumed with the next command.
With (for abort) the method is interrupted.
SAMPLE DATA
SAMPLE DATA 1: no current sample size
No sample size has been entered. Enter a sample size on the results page
(Softkey [Sample data]) and recalculate the determination (Softkeys
[ >> >>] and [Recalculate]).
SAMPLE SILO
SAMPLE SILO 1: sample silo full
SAMPLE SILO 2: too few empty lines in the sample silo
No further lines can be created in the sample silo. Lines without entries
(gray background) are regarded as being reserved. Delete any lines not re-
quired. If necessary, a sample silo which is full can be stored and the sam-
ple data it contains can be loaded later with the complete sample silo.