12.1 Error messages
796 Titroprocessor, Instructions for use
244
244
MALLOC
MALLOC 1:
last powerdown left inconsist. status. Continue anyway?
Switching off the instrument has caused memory conflicts. It is recom-
mended that the error message is answered with
tialization of the user memory area is carried out, see error message
DETSTORE 4, page 238. If the error message is answered with
troprocessor starts up without taking any further measures. However, it must
be expected that the instrument may behave in an irregular manner be-
cause of the memory conflict mentioned above.
MALLOC 3:
out of user memory
The working memory and the internal memory cannot accept any more
data. Delete determinations or methods which are no longer required from
these memories.
MALLOC 4:
out of system memory
The internal system memory is overloaded by printing jobs or other internal
functions. Terminate any printing jobs with the key combination
MALLOC 5:
reallocation of user memory failed
A serious fault has occurred during internal memory operations. Switch the
instrument off and then on again. If this error message appears again then a
total initialization of the whole memory is necessary which will delete all data
in the Titroprocessor. In order to trigger this switch the instrument on with
the space bar pressed down.
MALLOC 6:
reallocation of system memory failed
A serious fault has occurred during internal memory operations. Switch the
instrument off and then on again. If this error message appears again then a
total initialization of the whole memory is necessary which will delete all data
in the Titroprocessor. In order to trigger this switch the instrument on with
the space bar pressed down.
MAN_CAL
MAN_CAL 1:
invalid measured value
The measured value is not valid. Check the sensor and the buffer solution.
MAN_CAL 2:
temperature value invalid
The value given by the temperature sensor is invalid. Check it.
MAN_CAL 3:
value not in buffer table
A buffer value has been found which does not match any buffer solution of
the given manufacturer or type. Check the buffer parameters and the buffer
solution.