Metrohm 796 Titroprocessor User Manual
Page 290
13.5 Validation / GLP
796 Titroprocessor, Instructions for use
284
284
The method must contain the following results from the calculation formula:
•
Result for the statistical calculations
RT1=…
•
Theoretical result as numerical value
RT2=…
•
Endpoint which is to be used for the calculation of the systematic error
RE=EPx
(x=0…9)
Example of a titer determination for hydrochloric acid with TRIS as sample
substance.
ЪДДД * Formula ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ # Comment Formula (R$$, C##, X$$, EP# ...) Unit Dec Type ³
³ ДД ДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДД ДДД ДДДДДД ³
³ * 1 Titer RT1=SS*1000/0.1/121.14/EP1 4 stat. ³
³ * 2 Titer(theo) RT2=1.0000 4 stat. ³
³ * 3 Consumpt. RE=EP1 mL 3 stat. ³
The method must be stored.
13.5.4 Preparing for a validation
Call up the main page and press the
ЪДДД GLP ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ Method Validations Date Interval Result ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ 1. HCl Validation method ³
³ 2. ³
³ 3. ³
³ 4. ³
³ 5. ³
³ ³
³ ³
³ ³
³ ³
³ Instrument ident. : 01106 ³
³ Last servicing : 1999-11-25 ³
³ done by : B. Miller ³
³ Program version : 5.796.0010 ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ
ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿
³ ³ ³Delete ³ ³Define ³Extend.³Start ³ ³ ³Sample ³Sensors³
³ ³ ³method ³ ³method ³info ³valid. ³ ³ ³silo ³ ³
АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ
Five different validation methods can be defined for different purposes. Se-
lect a line and press the [Define method] softkey.